1.1 Power Supply

1.2 Gas input and output

1.3 Setting of the flow rate (controller only)

1. SETTING UP MASSFLOW GAS FLOW CONTROLLER / METER

The flow rate ranges depend on the MASSFLOW model and are as follows:

MASSFLOW 5000 (controller or meter) (Art. No.: 8071 or 8071-s)0.00 to 5.00 l/min in 0.01 l/min steps
MASSFLOW 500 (controller or meter) (Art. No.: 8072 or 8072-s)0 to 500 ml/min in 1 ml/min steps
MASSFLOW 500 hs (high sensitivity)(meter only) (Art. No.: 8072-hs)0 to 99.9 ml/min in 0.1 ml/min steps & 100 to 500 ml/min in 1 ml/min steps (auto-range)

 

1.1 Power Supply

When used together with the LAMBDA MINIFOR laboratory bioreactor the MASSFLOW gas flow controller or meter is connected with the corresponding 8-pole cable to the “PUMP”- socket at the rear of the MINIFOR laboratory fermenter-bioreactor. The other side of the cable is plugged into the REMOTE-socket at the rear of the MASSFLOW.

Figure 1.1-1 One end of the 8-pole cable should be connected to the “PUMP”- socket at the rear of the MINIFOR laboratory fermenter-bioreactor control unit.
Figure 1.1-2 Other end of the 8-pole cable should be connected to the REMOTE-socket at the rear of the MASSFLOW gas flow measurement and controller.

When used independently of the MINIFOR laboratory bioreactor-fermentor, a universal plug-in power supply (100-240 V AC/50-60 Hz, 12 VDC, 24 W) is used (art. no. 4821).

When the MASSFLOW is connected to the power supply, all LEDs and the display are lighted shortly. This allows a function control of all signal elements.

Figure 1.1-3 The 5-pole connector of the universal plug-in power supply (100-240 V AC/50-60 Hz, 12 VDC, 24 W) cable.
Figure 1.1-4 Plug the connector of power supply cable into the “POWER” socket (12 V) at the rear of the MASSFLOW gas flow controller or meter and secure it.

 

1.2 Gas input and output

The maximal gas pressure is 0.2 MPa (2 atm or 30 psig). Higher pressures will damage the instrument!

Fix the other tubing onto the gas OUT nozzle. Open the gas supply. The regulating valve is closed and no gas comes out from the output.

Figure 1.2-1 Connect the gas tubing (inside diameter about 5.5 to 6 mm) to the gas IN nozzle and secured it in place with appropriate clamps.

 

1.3 Setting of the flow rate (controller only)

Figure 1.3-1 Press the SET button. The corresponding LED will light and the display shows the pre-set value of flow rate.
Figure 1.3-2 Set a desired flow rate with the buttons Λ Λ Λ under the LED display.
Figure 1.3-3 Press the SET button again. The LED is switched off and the display shows the actual flow rate. It should indicate 0.0 l/min.
Figure 1.3-4 Press the ON/OFF button. After about 10 seconds (about 30 seconds for low flow rates), the flow rate will progressively attain the pre-set value.